Chapter 4 - Operational Procedures -
AS-410M Auto Slide Preparation System
99
When only the Cut Speed Offset is ON
The cut speed changes according to the cut
Force. The humidifying time which is set in
the selected sectioning parameter is applied.
When only the Humidify Time Offset is ON
The humidifying time changes according to
the measured cut forces. The sectioning speed
set in the selected sectioning parameter
is applied.
Cut Speed Offset & Humidify Time Offset
are OFF
Sectioning and humidification are conducted
based on the selected parameter settings.
The measured cut force is not referred to the
sectioning operation.
Cut Force monitor
The monitor displays the measured result of
cut force.
CUT FORCE SETTING screen
( Cut Speed & Humidify Time
Offset functions disabled )
Cut Force Monitor
measured result displayed
CUT FORCE SETTING screen
(Cut Speed Offset function enabled)
CUT FORCE SETTING screen
(Humidify Time Offset function enabled)

Table of Contents