Chapter 4 - Operational Procedures -
AS-410M Auto Slide Preparation System
99
<
When only the Cut Speed Offset is ON
>
The cut speed changes according to the cut
Force. The humidifying time which is set in
the selected sectioning parameter is applied.
<
When only the Humidify Time Offset is ON
>
The humidifying time changes according to
the measured cut forces. The sectioning speed
set in the selected sectioning parameter
is applied.
<
Cut Speed Offset & Humidify Time Offset
are OFF
>
Sectioning and humidification are conducted
based on the selected parameter settings.
The measured cut force is not referred to the
sectioning operation.
<
Cut Force monitor
>
The monitor displays the measured result of
cut force.
( Cut Speed & Humidify Time
Offset functions disabled )
(
measured result displayed
)
(Cut Speed Offset function enabled)
(Humidify Time Offset function enabled)