EasyManua.ls Logo

Dräger Infinity Vista - Recording Waveforms; Timed Recordings

Dräger Infinity Vista
280 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-6 Infinity Vista VF6
7 RECORDINGS
Recording Waveforms
To print real-time monitoring data, two types of recordings are available:
z Timed Recordings.
z Continuous Recordings.
For both recording types, the recording speed is 25 mm/s.
Timed and continuous recordings include up to two waveforms. If one of the
waveforms is hidden by a menu, you can still request a recording. If a cascaded
waveform is displayed, only the waveform from the first channel prints out.
Timed Recordings
Timed Recordings print a total of 20 seconds of patient data, of which about 10
seconds occurred prior to the recording request (Delay), and about 10 seconds after
the recording request.
If no recorder/printer is available, the monitor stores the recording (see Stored
Recordings, below). You can cancel the storage within 5 seconds of the initial
recording request.
The monitor cancels a timed recording if you modify the display, size or scale of a
waveform while the recording is ongoing.
STEPS: Starting a Timed Recording
1. Press the monitor’s Record fixed key.
2. To cancel the recording, press the
Record key again or the Stop key on the
recorder while printing is in progress.
NOTE: Timed and continuous recordings cannot be started while the Alarm Limits
table, trends or diagnostic logs are displayed. To start recording, exit those displays
before requesting a recording.