5-10 Infinity Vista VF6
5 ALARMS AND MESSAGES
All Alarms OFF Key
Press the All Alarms OFF fixed key to suspend all alarm indications for 3 minutes.
During an alarm suspension, the message banner All Alarms OFF appears on top of
the waveform area.
During the 3-minute alarm suspension:
z All currently annunciated alarm tones cease.
z All parameter boxes return to normal colors.
z All alarm messages are removed from the message area.
z All new alarms are blocked.
z Currently activated alarms, including latched alarms, are acknowledged.
z When on the network, the message banner All Alarms OFF appears in the
bedside monitor’s remote display on the central station and in the
ClusterView.
NOTE: When starting the monitor, coming out of Standby, or answering “Yes” to the
New Patient prompt, alarms are suspended for 3 minutes or until you press the All
Alarms OFF key to deactivate the alarm suspension.