ALARM AND STATUS MESSAGES
VF6 Infinity Vista 3-7
Alarm and Status Messages
When the monitor is connected to the network, network messages alert you of network
operating conditions. Some messages display only once (i.e., Remote Limit Change),
while others appear alternately until the condition has been resolved (i.e. alarm messages
such as BED 200:ECG Leads Off). A network alarm error creates an error tone, while
status messages do not.
Screen Message Condition
Selected Bed Label
Currently in Use
The user selects a care unit and bed label for a wireless monitor,
but the bed label is already in use.
Duplicate Address Another network device has been programmed with a conflicting
identification. The bedside monitor is treated as being offline.
Network Alarm Error An interruption in the network communication has been detected
while an alarm is active; the speaker volume at the bedside
monitor has been increased to its maximum level.
Not Monitored by Central A wireless monitor is transmitting to a central station, but is not
assigned a viewport on the central station’s cluster screen.
Offline The monitor is not connected to the network or the network is not
configured correctly.
A wireless monitor has traveled out of range.
The wireless card is not installed properly.
Remote Limit Change Alarm settings (on/off, alarm limits, alarm recording, including
ARR and ST) have been changed at the central station.
Remote Relearn Relearning of the patient’s ECG or respiration pattern has been
initiated at the central station.
Silence by Remote Alarm has been silenced at the central station.