RESPIRATION MONITORING SETTINGS
VF6 Infinity Vista 11-11
Relearning a Patient’s Respiration Pattern
When respiration monitoring is enabled and set to Auto, the monitor takes about one
minute to automatically learn the patient's respiration pattern. You can initiate a new
learning phase at any time. When respiration monitoring is enabled and set to
Manual,
you must initiate the learning of the respiration pattern manually. Whenever you
change or reposition electrodes or if you notice significant changes in the patient's
breathing pattern, repeat the learning process.
STEPS: Relearning a Patient’s Respiration Pattern
1. Click on the Rsp parameter box.
2. Click on
Relearn.
During the learning phase, the monitor displays the message Rsp Relearning at the
bottom of the screen and the
LRN label in the Rsp parameter box. During the learning
phase, the patient should rest as quietly as possible and breathe normally.