3-2 Infinity Vista VF6
3 NETWORK APPLICATIONS
Overview
The Infinity network provides communication links between the bedside monitor and
other network devices such as other monitors, a central station, recorders and laser
printers. The network allows you to monitor your patient at a central location away
from the bedside, and to monitor many patients at once. Within the Infinity network
you can:
z Receive alarm messages at the local bedside monitor from other devices
within the network.
z Print recordings on network recorders and laser printers.
z Transfer patients between monitors in the network.
z View patient data of up to 16 patients at a dedicated central station.
z Admit patients at a central station.
z Control alarms from a central station.
z Initiate a Relearn of a patient’s ECG and respiration pattern from a central
station.
z Set arrhythmia parameters from a central station.
z Collect diagnostic logs at a central station.
Basic Network Components
Infinity Vista with
Wireless LAN PC Card
Wireless LAN PC Card
Access Point