DEFAULT S ETTINGS
VF6 Infinity Vista C-3
Adult
Settings
Pediatric
Settings
Neonatal
Settings
After patient discharge,
settings return to...?
ECG
Alarm On On On Saved Setup
Alarm Limits
(upper/lower)
120/45 150/50 180/80 Saved Setup
Alarm Recording Off Off Off Saved Setup
Tone Source ECG ECG ECG Previous Patient
Tone Volume Off Off Off Previous Patient
Pacer Detect Off Off N/A Saved Setup
QRS Marks Off Off Off Factory
ECG Processing ECG 1&2 ECG 1&2 ECG 1 Previous Patient
ECG Leads 5-Lead 5-Lead 3-Lead Previous Patient
Arrhythmia*
ASY, VF Alarms Always On Always On n/a Previous Patient
VT, RUN, AIVR,
SVT, CPT, BGM
Alarm
BRDY, TACH,
PAUS, ARTF Alarm
On
Off
On
Off
n/a
n/a
Saved Setup
ASY and VF Alarm
Recording
Other Events
Record
Off
Record
Off
n/a
n/a
Saved Setup
Arrhythmia
Monitoring
Basic Basic n/a Saved Setup
*Not applicable for neonatal monitoring.