INSTRUCTIONS FOR USE
VF6 Infinity Vista iii
Infinity Vista Software Release Notes
Software Version VF6
z Wireless network operation requires special configurations of the monitoring
network (a service function). If you experience problems with wireless
network operations, contact your Service personnel.
z When moving and assigning a wireless monitor to a different central station,
the original central station may emit a brief network error tone and display an
Offline message instead of the message Bed Disconnected. However, there is
no disruption of network monitoring and the Offline message clears as soon as
you assign a new bedside monitor to the central viewport.
z When you change the units of measure at the bedside monitor and the central
station is showing the monitor’s bed view, you must first exit the central bed
view, before the change of units appears at the central station.
z For network and card data transfer:
— Occasionally, after a data transfer from an Infinity Vista XL/Gamma X XL
monitor, you may see three or four ST trends instead of the two ST leads
monitored by the Infinity Vista monitor.
z For network data transfer only:
— After a network transfer of telemetry data to an Infinity Vista monitor, ST
trend points may appear two minutes apart.
z NBP parameter values transferred from an Infinity Vista monitor to an
Infinity Vista XL/Gamma X XL monitor are displayed in the trend graphs
rather than in the trend tables of the destination monitor.
z When power-cycling the monitor or admitting a new patient, saved
monitoring settings may occasionally return to default settings. Check
monitoring settings after these events.
z Occasionally, the ECG waveform is not displayed in the second waveform
channel, when you assign SpO
2
to the first waveform channel. In this case,
click on the second waveform channel and select the desired ECG lead again.
z If two wireless bedside monitors are given the same bed label or IP address
during manual network configuration (a Biomed function), their connection to
the central station alternates between online/offline. In this case, change one
monitor’s bed label in the Admit menu or contact your Biomed technician to
change one monitor’s IP address.
z When you change parameter colors at a networked bedside monitor, the
monitor temporarily goes offline at the central station in order for the color
change to take effect.