RECORDING WAVEFORMS
VF6 Infinity Vista 7-7
Continuous Recordings
Continuous recordings provide an up-to-the-second printout of patient data. This
mode is useful when more than 20 seconds of data must be printed during critical,
short-term applications.
STEPS: Starting a Continuous Recording
1. Press the Record fixed key for at least two seconds. Two beeps indicate when
the recording starts.
2. To cancel the recording, press the
Record fixed key again or the Stop key of
the recorder while printing is in progress.
Continuous recordings cannot be stored. If you attempt to store a continuous
recording, the monitor stores it as a timed recording with 20 seconds of patient data.
NOTE:
z You can request a continuous recording during an alarm, timed, trend, and
diagnostic log recording. To do so, first stop the recording in progress with the
Record fixed key on the monitor, then press and hold the same key to start the
continuous recording.
z Continuous OCRG recordings are not supported.
NOTE:
z If the user requests a continuous recording at the central station while a timed
recording is in progress, the timed recording is cancelled. A continuous
recording can only be printed out if two recorders are available. If only one
recorder is available, the continuous recording request is converted to a timed
recording, then stored and printed when the recorder becomes available.
z An uninterrupted continuous recording on a laser printer consists of up to 120
pages.