ST ALARMS
VF6 Infinity Vista 10-7
ST Alarms
Set ST alarm limits on the Alarm Limits table (see the chapter Alarms and Messages).
The default ST segment deviation alarm limits are +1.0 mm (or +0.10mV) for the
upper limit and -1.0 mm (or -0.10 mV) for the lower limit. However, adjustments to
the upper and lower alarm limit settings must be made based on careful clinical
evaluation. You can set the upper and lower limits within the following range:
z -15.0 mm to +15.0 mm, in 0.1 mm increments (-1.5 mV to +1.5 mV, in 0.01
mV increments, when using millivolts.)
NOTE: Always check the upper and lower ST alarm limits before starting ST
monitoring.