RELEARNING A PATIENT’S ECG
VF6 Infinity Vista 9-7
Arrhythmia Alarm Recordings
STEPS: Turning Alarm Recordings On/Off
1. Call up the Arrhythmia Setup table (see above).
2. Click on the recording setting in the
Record column of the desired arrhythmia
parameter.
3. Dial in the desired setting (
Record, Store, Str/Rec, or OFF).
Relearning a Patient’s ECG
It is advisable to initiate a relearning under one or more of the following conditions:
z Eight hours have passed since the last learning of a reference complex.
z You repositioned the electrodes.
z You observe clinically questionable arrhythmia calls.
z There have been significant changes in the patient’s ECG.
z The message ARR Relearn appears in the message area.
The monitor automatically relearns in the following cases:
z When you turn arrhythmia monitoring on or change between basic and full
arrhythmia monitoring.
z When you change the displayed ECG leads while arrhythmia monitoring is
on.
NOTE:
z You cannot turn alarm recordings for asystole (ASY) and ventricular fibrillation
(VF) off. For these events, only the settings
Record and Str/Rec are available.
z See the chapter Recordings for a description of alarm and stored recordings and
how to view, save, print, or delete stored recordings via the Event Recall screen.