9-8 Infinity Vista VF6
9 ARRHYTHMIA
STEPS: Relearning the Patient’s ECG
1. Make sure that the ECG lead wires are properly connected and that the ECG
displayed seems normal for this patient.
2. Click on the
HR parameter box.
3. Click on
More....
4. Click on
Relearn.
During the learning phase, LRN appears in the HR parameter box and the message
Arrhythmia Relearning in the message area at the bottom of the screen. Once learning
is complete, date and time of the learning phase are displayed in the HR Setup menu
next to the Relearn selection.
NOTE: You cannot initiate a relearning, if arrhythmia monitoring is turned off.