ALARM S ETTINGS
VF6 Infinity Vista 5-5
Setting Alarm Limits
You can set alarm limits for each parameter individually, or you can use the AutoSet
function to set alarm limits for all parameters at once based on the current parameter
values.
STEPS: Setting alarm limits individually
1. Press the Alarm Limits fixed key.
2. Scroll to the
Upper column for the desired
parameter and click the knob.
3. Dial in the desired upper limit and click the knob.
4. Scroll to the
Lower column and click the knob.
5. Dial in the desired lower limit and click the knob.
STEPS: Using the AutoSet function
1. Press the Alarm Limits fixed key.
2. Scroll to
AutoSet. The column AutoSet shows
the upper and lower limits that are automatically
calculated for each parameter based on current
parameter values.
3. To accept these limits for all parameters, click on
AutoSet. The automatically calculated limits are
written into the Upper and Lower columns.
NOTE:
z The AutoSet function sets alarm limits for all parameters at once, even for those
not displayed on the current page of the Alarm Limits table. Exception: There is
no AutoSet function for PVC.
z If a parameter value is not currently available (e.g. during a lead-off condition),
the monitor does not adjust the alarm limit for that parameter when you click on
AutoSet.
z After using the AutoSet function, you can change individual alarm limits at any
time.
z Due to slight rounding differences in values, the alarm limits display for kPa
values at the central station does not match the alarm limits display for kPa
values at the Infinity Vista monitor.