15-2 Infinity Vista VF6
15 TEMPERATURE
Overview
The monitor supports the use of temperature probes for oral, rectal and axillary
applications. The monitor accepts one probe at a time and displays absolute
temperature values in degrees Celsius (
o
C) or in degrees Fahrenheit (
o
F).
To calculate a temperature value, the monitor averages the temperature signal over
several seconds. The temperature label and measurement value appear above the first
waveform channel. There is no waveform display for temperature.
Set T alarms on the Alarm Limits table (see the chapter Alarms and Messages). If you
turn T alarms off, a crossed bell symbol appears next to the temperature value above
the first waveform channel.
Temperature Probes
A wide selection of reusable and disposable thermistor probes is available. Use only
Dräger-approved probes (see the appendix Options and Accessories); other probes are
not recommended and may produce inaccurate measurement results.
Placing the Probe
Rectal Probes
To place the probe, follow the clinical techniques of your hospital. We also suggest the
following:
1. Mark the insertion depth (2 to 4 inches) with a rubber ring or tape.
2. Insert the probe through the rectum into the colon and tape the cable in place.
NOTE:
z Selecting the unit of measure is a password-protected function. See your
Biomed for details.
z When the monitor uses the export protocol to communicate with an external
device, the temperature values in degrees Celsius (
o
C) are transmitted as whole
numbers without a decimal point on the external device. For example, a
temperature value of 37.6 degrees is transmitted as 376.
NOTE: Cover reusable probes with a protective rubber cover.