C-16 Infinity Vista VF6
C DEFAULT SETTINGS AND BIOMEDICAL SUPPORT
NBP/SpO
2
Interlock
When NBP/SpO
2
Interlock is ON, SpO
2
alarms are disabled during an NBP
measurement in order to avoid nuisance SpO
2
alarms caused by impeded circulation
in the limb.
Turn NBP/SpO
2
Interlock on as follows:
1. Press the
Menu fixed key.
2. Click on
Monitor Setup.
3. Click on
Biomed.
4. Enter the Biomed password and click the knob again.
5. Click on
Monitor Setup.
6. Click on
NBP/SpO
2
Interlock.
7. Turn the function
ON or OFF.
ASY/VF Alarms
Set alarms for asystole and ventricular fibrillation as either Always ON or Follow HR.
If you select Follow HR, ASY and VF alarm settings follow those settings selected for
heart rate alarms in the Alarm Limits table. If you select Always ON, the monitor
always alarms for asystole and ventricular fibrillation, even if HR alarms and
arrhythmia monitoring are turned off.
Set ASY/VF Alarms as follows:
1. Press the
Menu fixed key.
2. Click on
Monitor Setup.
3. Click on
Biomed.
4. Enter the Biomed password and click the knob again.
5. Click on
Monitor Setup.
6. Click on
ASY/VF Alarms.
7. Select Always ON or Follow HR.