DEFAULT S ETTINGS
VF6 Infinity Vista C-5
PLS Alarm Limits
(upper/lower)
120/45 150/50 170/80 Saved Setup
Temperature
Alarm Off Off Off Saved Setup
Alarm Limits
(upper/lower)
39/34 °C
102.0/93.2
°F
39/34 °C
102.0/93.2
°F
39/34 °C
102.0/93.2
°F
Saved Setup
Alarm Recording Off Off Off Saved Setup
Units °C °C °C Previous Patient
NBP
Alarm Off Off Off Saved Setup
Alarm Limits
(upper/lower)
• Systolic
•Mean
• Diastolic
• 160/90
• 125/60
• 110/50
• 160/90
• 125/60
• 110/50
• 80/50
• 70/40
• 60/25
Saved Setup
Alarm Recording Off Off Off Saved Setup
Interval Mode Off Off Off Saved Setup
Calibration Mode Off Off Off Previous Patient
Inflation Mode Adult:270 Ped.: 180 Neo.: 140 Previous Patient
Measurement Tone Off Off Off Saved Setup
IBP1, IBP2
Alarm Off Off Off Saved Setup
Alarm Limits
(upper/lower)
• Systolic
•Mean
• Diastolic
• 160/90
• 125/60
• 110/50
• 160/90
• 125/60
• 110/50
• 120/50
• 85/40
• 80/35
Saved Setup
Alarm Recording Off Off Off Saved Setup
Cal. Factor 100 100 100 Previous Patient
Manometer Cal. 100 100 100 Previous Patient
Adult
Settings
Pediatric
Settings
Neonatal
Settings
After patient discharge,
settings return to...?