D-8 Infinity Vista VF6
D TECHNICAL DATA
Pacer Pulse Detection Level Amplitude: ± 5 to ± 700 mV
Width: 0.1 to 2.0 ms
Separation between pulses: t30 ms
Recharge time constant: 4 to 100 ms
Over/Undershoot: 0.025 a
p
2 mV maximum
Lead Off Sensing Current <50 nA DC from LL, LA, RA and V Leads
Arrhythmia
Measurement Range 0 to 300 beats/minute
Display Resolution 1 beat/minute
Accuracy ± 5 beats/minute or ± 10% of the rate, whichever is greater
Response Time < 4 seconds
ST Segment Analysis
Measurement Range -15 mm to +15 mm (-1.5mV to +1.5 mV)
Accuracy ±1mm/0.1mV RTI
Isoelectric Point Range From start of the averaged ST complex to Fiducial Point, in
increments of 4 ms
ST Point Range From Fiducial Point to end of the averaged ST complex, in
increments of 4 ms
Display Resolution 0.1 mm/0.01mV (10PV)
Response Time 15 seconds
Respiration
Method Impedance Pneumography
Sensing Electrodes RA and LL (Lead II)
Number of Channels 1
Sweep Speed 6.25 mm/s