D-10 Infinity Vista VF6
D TECHNICAL DATA
Measuring accuracy, Adult mode
(1)
:
SpO
2
:
0 to 69% not specified
70 to 100% sensor-specific as follows:
Nellcor :
(2,3)
D-25/D-25L, D-20, I-20, N-25, OxiMAX MAX-A, -P, -I, -N, -AL ................................... ±2
Nellcor :
DS100A ...................................................................................................................... ±3
Masimo:
(2,3)
LNOPADT, LNOPPED, LNOPNEO, LNOPNEO SS,LNOP-YI .................................... ±2
Masimo:
LNOP-DCI, LNOP-DCIP, NR125 ................................................................................ ±2
EAR ......................................................................................................................... ±3.5
Pulse Rate: ................................................................±3 beats/min or ±3% (whichever is greater)
Measuring accuracy, Neonatal mode
(1, 2, 3, 4)
:
SpO
2
:
0 to 69% not specified
70 to 100% sensor-specific as follows:
Nellcor:
N-25 ...........................................................................................................................3
Masimo:
LNOPNEO, LNOPNEO SS, LNOP-YI......................................................................... ±3
Pulse Rate: ................................................................±3 beats/min or ±3% (whichever is greater)
Notes:
1) SpO
2
accuracy specification is at 1 SD (standard deviation) and is expressed as ± "X" digits
(absolute) for the established range.
2) Accuracy of saturation measurements on neonates is increased by ±1 digit as compared to
accuracy on adult patients to account for the theoretical effect on oximeter measurements of
fetal hemoglobin in neonatal blood.
SpO
2
Alarms High: Adjustable, 20 to 100%
Low: Adjustable, 20 to 100%
Defaults:
Adult and Pediatric: 100-90
Neonatal: 100-90
Nominal Wavelength Red: 660 nm
IR: 910 nm
Power Red: 3 mW (max.)
IR: 4 mW (max.)
Note: LED drive is current limited by hardware mechanisms.
Degree of Protection Against
Electric Shock
Type CF
Defibrillation Protection In accordance with IEC 601-1A2
Pulse Oximetry (SpO
2
) (continued)