D-18 Infinity Vista VF6
D TECHNICAL DATA
Recommended separation distances between portable and mobile RF communications equipment and
the equipment
Rated maximum output
power of transmitter W
Separation distance according to frequency of transmitterin meters
150 kHz – 80 MHz
d=1.2/V1]P
80 MHz to 800MHz
d=1.2/V1]P
800 MHz to 2.5 GHz
d=2.3P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and
reflection from structures, objects and people.