5-14 Infinity Vista VF6
5 ALARMS AND MESSAGES
Heart Rate, Arrhythmia, ST Segment Analysis
Message Possible Cause Suggested Action
Message: Arrhythmia
Relearning
Parameter value: LRN
The monitor is learning the
patient’s normal QRS complex to
use as reference.
Wait until the message
disappears from the screen.
Message: Asystole
Parameter value: ASY
No QRS detection for the last 4
seconds.
Heart rate is below 15 beats/
minute.
Observe the patient and treat if
clinically indicated.
Select a lead with at least an
amplitude of 0.5 mV.
Select another lead to meet the
above criterion.
Reposition or change the
electrodes if the amplitude is still
low.
Use good skin preparation.
Message: RL Lead Off
*
HR parameter value:
<value>
ST<lead1>/ST<lead2>
parameter value:
<blank>
Message: V Lead Off *
Message: V+ Lead Off
Message: RA Lead Off
Message: LL Lead Off
Message: LA Lead Off
Message: ECG Leads
Off
HR parameter value:
***
ST<lead1>/ST<lead2>
parameter value:
<blank>
The monitor has detected a lead-off
condition for the currently
processed lead. The cause could
be one of the following:
Unplugged cable
Broken cable
Loose lead wire
Faulty lead wire
Dried out gel on the electrode
Wrong cable type selection.
The M
ULTIMED/NEOMED pod may
be defective.
Check the patient cable and lead
wires carefully.
Check the M
ULTIMED/NEOMED
pod. Replace pod if defective.
Replace any cable or lead wire
that is suspect.
Reapply gel on the electrode or
change the electrode (if
disposable).
Select another ECG lead for
processing if the electrode or lead
cannot be replaced.
* For all the electrode disconnected messages (V, V+ RA, LL, LA, RL) the parameter box shows
*** if this is the lead currently being monitored. If another valid lead is being monitored (i.e. ECG
1&2) then the parameter value for HR, ST<lead1> and ST<lead2> is <value>
Message: ARR
Cannot Learn Lead
<lead>
Parameter value:
<value>
Lead not connected. Check lead connections