Messages
Message
Location
PossiblecausesSuggestedactions
P1zeroingfailedtoP2
zeroingfailed
param.
Thechannelhasnotbeenzeroed
successfully.
Repeatthezeroing.
PA1syshigh/PA1
syslowtoPA2sys
high/PA2syslow
PA1meanhigh/PA
1meanlowtoPA
2meanhigh/PA2
meanlow
PA1diahigh/PA1dia
lowtoPA2diahigh/
PA2dialow
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
PR(Art1)hightoPR
(Art2)high/PR(Art1)
lowtoPR(Art2)low
PR(Fem1)hightoPR
(Fem2)high/PR(Fem
1)lowtoPR(Fem2)
low
PR(UAC1)hightoPR
(UAC2)high/PR(UAC
1)lowtoPR(UAC2)
low
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
Pressuredevice
failure:Callservice
al.areaFaultyCARESCAPEPressure.Contactqualiedservice
personnel.
Pressuredevice
initializing
al.areaNewsoftwareisbeing
downloadedtotheCARESCAPE
Pressure.
Wait.Themessagewill
clearwhenthedownloadis
complete.
Pressuredevice
overheated.Shutting
itdown.
al.areaCARESCAPEONEturnsoffthe
CARESCAPEPressurebecauseits
surfacetemperatureistoohigh.
ChecktheCARESCAPE
Pressureplacementand
ensurethatitisnotcovered
byanythingthatmightlead
tooverheating.
Iftheproblempersists,
replacetheCARESCAPE
Pressure.
Pressure
measurement
removed
al.areaCARESCAPEPressurehasbeen
removedoractivepressure
channelbecomesinactive.
Checkallconnectionsand
reconnectasrequired.
PressureSensed
param.
Pressurehasbeensensedduring
zeroing.
Opentheventingstopcockto
air.
Re-zero.
RAP1meanhigh/
RAP1meanlowto
RAP2meanhigh/
RAP2meanlow
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
2094480-001CARESCAPEONE223