Alarmspecications
AlarmParameterAlarmdelay
MoveSpO2devicetoacooler
location
SpO
2
6s
NIBPcuffloose
NIBP
22s(average)(22srange)
NIBPcuffocclusion
Cuffocclusion
NIBP<5s
NIBPautostopped
NIBPSTATstopped
NIBP<5s
NIBPoverrange
NIBPunderrange
NIBP<5s
Noise
NoisyECG
ECG
27s(average)(25to27srange)
Pressuredevicefailure:Call
service
Invasivepressures
7s
Pressuredeviceinitializing
Invasivepressures
6s
Pressuredeviceoverheated.
Shuttingitdown
Invasivepressures
10s
Pressuremeasurement
removed
Invasivepressures
<5s
Resp(Imped.)measurement
paused
Measurementpaused
Impedancerespiration50s(average)(49to51srange)forall
softwarepackagesexceptNICU.
NICUsoftwarepackage:18s(average)
(17to20srange)
Samplelinedisconnected
Linedisconnected
CO
2
42s(average)(41to42srange)
SpO2devicefailure:Call
service
SpO
2
6s
SpO2deviceoverheated
Shuttingdown
SpO
2
11s(average)(9to12srange)
SpO2faultyprobe
FaultyadhesiveSpO2probe
SpO
2
7s
SpO2measurementremoved
SpO
2
<5s
Tempdevicefailure:Call
service
Temperature
8s
Tempdeviceoverheated.
Shuttingdown
Temperature
8s
Tempmeasurementremoved
Temperature<5s
272CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017