Alarmspecications
AlarmParameterAlarmdelay
Deviceoverheating
MoveTempdevicetoacooler
location
Temperature7s
ECGdevicefailure:Callservice
ECG
7s
ECGdeviceoverheated.
Shuttingdown
ECG10s
ECGmeasurementsremoved
ECG
<5s
IdenticalECGdevice
ECG
<5s
Identicalpressuredevices
Invasivepressures
<5s
IdenticalSpO2device
SpO
2
<5s
IdenticalTempdevice
Temperature<5s
Incompatibleprobe
IncompatibleSpO2probe
IncompatibleadhesiveSpO2
probe
SpO
2
7s
IncompatibleSpO2device
SpO
2
8s
[Invasivepressurechannellabel]
Xsensordisconnected,where
[Invasivepressurechannellabel]
=Art,CPP,CVP,Fem,FemV,ICP,
LAP,P,PA,RAP,RVP,UAC,or
UVC,andX=invasivepressure
channelnumber1to2.
Sensor
Invasivepressures
6s
[Invasivepressurechannel
label]Xsensorfailed,where
[Invasivepressurechannellabel]
=Art,CPP,CVP,Fem,FemV,ICP,
LAP,P,PA,RAP,RVP,UAC,or
UVC,andX=invasivepressure
channelnumber1to2.
Sensor
Invasivepressures
8s
IP'snotzeroed
Invasivepressures
301s(average)(300to301srange)
LA/Lleadoff
LL/Fleadoff
RA/Rleadoff
RL/Nleadoff
ECG
LA/Lleadoff,LL/Fleadoff:10s
RA/Rleadoff:11s(average)(10to11
srange)
RL/Nleadoff:16s(average)(15to17
srange)
Leadsoff
ECG
7s
●MoveCO2sensortoacooler
location
●Sensoroverheating
CO
2
<5s
2094480-001CARESCAPEONE271