GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 5344303-100, REVISION 3 VIVID P3 SERVICE MANUAL
8-30 Section 8-4 - Mechanicals
8-4-3 Rear Cover (FRU No. 5315117)
Purpose: This is a description on how to remove and replace the Rear Cover.
8-4-3-1 Tools
• Common Phillips screwdrivers
8-4-3-2 FRU BOM. This FRU Kit P/N:5315117 consists of following parts.
8-4-3-3 Needed Manpower
• 1 persons, 15 minutes
8-4-3-4 Preparations
• Shutdown the system and switch off the main Circuit Breaker at the bottom rear side of the system.
8-4-3-5 Removal Procedure
1.) Remove RH-side cover from the system. Refer section 8-4-1 on page 26
2) Remove LH-side cover from the system. Refer section 8-4-2 on page 28
3) Remove the two screws (A,B) RH- side from the system. Refer Figure 8-26 on page 8-31
Figure 8-24 Rear cover
PART NO DESCRIPTION
5310092 Rear_Cover_Lakshya
2334843
Fasteners- M4 X 6 ECO-FIX
5314208 Caution Label Big
Rear Cover