GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 55344303-100, REVISION 3 VIVID P3 SERVICE MANUAL
7-20 Section 7-6 - Troubleshooting
7-6-1-9 Trackball
Figure 7-16 Trackball
The trackball is low sensitive!
Remove the ball from the Trackball,
and clean it.
YES
NO
Doesn’t the trackball move
at all?
END
The trackball module is defective
and replace it in Key board.