GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 5344303-100, REVISION 3 VIVID P3 SERVICE MANUAL
8-86 Section 8-13 - Loading Base Image Software
Section 8-13
Loading Base Image Software
FRU part num for Base Image software.
5344308-3 Vivid P3 Application Software with Base Image for Expert.The software DVD is kept in the
Rear EMI cover of the system.
NOTE: While it is believed to be unnecessary, It would not hurt to disconnect the system from the
network and remove all transducers.
NOTE: To Enter BIOS Press F2 (Arrow) Key immediately after power on and enter BIOS Password for
LP3 as Lakshya.
NOTE: Please ensure AC adapter is connected during system upgrade!
1.) Before loading the Software, backup the database. For reference see 4-3-3-1 on page 4-5 and
after loading the Software, restore the backup database.
2.) Insert the disk labeled “System & Application Software” into the DVD drive.
3.) Properly turn off the scanner by momentarily pressing the Power On/Off Switch. Select “Full
Maintenance Reboot” from the System Exit menu.
4.) If the system will not shutdown normally, hold down the Power On/Off Switch until the light
turns from green to amber.
5.) Turn on the scanner. System will detect the DVD-RW automatically.
Figure 8-72 Shutdown Dialog Box