IRECTION 5344303-100, REVISION 3 VIVID P3 SERVICE MANUAL
Chapter 4 Functional Checks 4-5
4-3-3 Archiving and Loading Presets
NOTE: Always save presets before any software reload. This ensures the presets loaded after the software
reload are as up–to–date as possible.
All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an DVD-
R disk (or USB memory device) for reloading on the system.
4-3-3-1 Archiving Presets to an DVD-R Disk (or USB memory device)
1.) Insert an empty (blank) DVD-R disk into the DVD-RW.
2.) Access the Config Menu, and select System. The Backup sheet will be shown on the LCD display.
3.) Select the item to back up from the Backup section.
4.) Enter backup destination or browse through the disk to locate the destination.
5.) Select Backup now. The backup status for each item is displayed on the Result column.
NOTICE
Presets should NOT be saved on the same DVD-R disk (or USB memory device) as images. The
Archive Menu lists the images but does NOT list the presets stored on a DVD-R disk (or USB memory
device).