GE HEALTHCARE PROPRIETARY TO GE
D
IRECTION 5344303-100, REVISION 3 VIVID P3 SERVICE MANUAL
Chapter 10 Care & Maintenance 10-31
ELECTRICAL SAFETY
Final Check. All system covers are in place. System scans with all probes as expected.
Accepted by: ______________________________________________________________________
Electrical Test Performed
Max Value
Allowed
Value
Measured OK? Comments
Outlet (correct ground &wiring config.)
Type BF Applied Part Leakage Current
Limits- Probe
enclosure Source Leakage Current -
Chassis Leakage Current Limits
Peripheral 1 Leakage Current
Peripheral 2 Leakage Current
PROBES
Probe Number
(from previous page)
Max Value
Allowed
Max Value
Measured OK? Comments
Probe 1:
Probe 2:
Probe 3: