80
ENGLISH
Mode 1: Automac inaon for the pressure accuracy vericaon
At this mode, the monitor can acvate the inaon, so the pressure will increase automacally unl it exceeds the limit
value specied in table A. This pressure limit value depends on the paent type selecon as shown in table A:
Adult 240mmHg
Pediatric 200mmHg
Neonate 120mmHg
Table A
During the inaon, the Monitor will close the deang valve, and the pressure value will be shown during the process.
If there is no manual deaon operaon, the pressure will persist unl deaon by manual operaon, so it is necessary
to use a manual valve for doing adequate deaon in several steps to verify the pressure accuracy in the full scale of
measurement range.
Mode 2: Manual inaon for the pressure accuracy vericaon.
At this mode, the pressure should be increased manually by a pumping balloon, and the vericaon can be done by
applying dierent pressure value manually. If the increased pressure exceeds the given limit as shown in table B, the
Monitor will deate automacally because of over-pressure protecon.
Adult 300mmHg
Pediatric 240mmHg
Neonate 140mmHg
Table B
Aer the vericaon, do press the buon again to return to normal working mode, then connue other
operaon, or the NIBP key will be invalid.
Pressure accuracy vericaon must be operated by technician or equipment manager. Doctor or nurse is not
allowed to do the vericaon, it is very dangerous especially when the pressure cu is sll on paents.
Air Leakage Check
In order to avoid signicant error of blood pressure measurement or even no measurement result caused by air leakage
in the pneumac system including the cu during measuring, it is recommended to check if there is leak in the pneumac
system as well.
Please remove the cu from paent while performing the leakage check.