Operation - Calibration
Fusion4 MSC-L Part No.: 4418309_Rev09
5 - 266 Installation & Operation Manual
Honeywell
3. Select <OK> on the IR controller or LAD to select the calibration point.
The Calibration . Stream n . Edit Meter Profile . Wizard screen
appears when a calibration record is selected.
The following functions can be performed.
Add - Adds a new calibration point to the meter profile.
Delete - Deletes the selected calibration point.
Calibrate - Re-calibrates the selected calibration point.
Saves - Saves the meter profile (generates Calibration Log
record).
NOTE: While adding a new calibration point, the initial meter
factor is considered to be 1.