Operation - Calibration
Part No.: 4418309_Rev09 Fusion4 MSC-L
Honeywell Installation & Operation Manual 5 - 267
4. If the Add function is selected, then the Calibration . Arm n . Stream
n . Wizard screen appears, on which the preset volume must be
entered.
5. Enter the preset volume and then select <Confirm>.
The Calibration . Arm n . Stream n . Wizard screen appears,
on which the flowrate must be entered.
6. Enter the flowrate and then select <Confirm>.
The Calibration . Arm n . Stream n . Wizard screen appears,
on which the start delay time in seconds must be entered. A suitable
delay time must be entered so that you can move to another