Operation - Calibration
Fusion4 MSC-L Part No.: 4418309_Rev09
5 - 274 Installation & Operation Manual
Honeywell
4. On the Calibration . Stream n . Wizard 2/6 screen, enter the start
delay time in seconds and then select <Confirm>.
A suitable delay time should be entered so that you can move to
another place (for example, to observe the result) until the process
continues.
The Calibration . Stream n . Wizard 3/6 screen appears with the
preset volume, delay time in seconds, and the progress status.
5. Check the Progress bar to monitor the calibration progress.
6. Select <OK> on the IR controller or LAD.
The countdown of the delay time starts. After the expiration of the
delay time, the injection process starts.