Serial interface RS485 
Connections  include, for  example,  other  measurement devices,  devices  for final processing  and 
evaluation,  PLC  systems,  and  so  on.  Cable  type  refer  to    5.4.2.Cables  (power  supply  / 
communication). 
There are 2 interfaces in the connection box (TB4-terminals 1 to 3 for CH1 and TB4-terminals 4 to 
6 for CH2, details see figure 5.6). The connection varies depending on the device (see dotted cable 
in the following example). The ground (0 V) is either connected  directly  to  the  device/system or 
connected to GND.  
Pull-up / pull-down resistors should be installed at the cable end near the data evaluation unit to 
generate the neutral potential. A 470 Ω resistor must be connected between RSA and the positive 
supply voltage of the connected data evaluation unit. A further 470 Ω resistor must be connected 
between RSB and GND. For cable lengths over 200 m ≙ 656ft, additional bus connection resistors 
of 120 Ω each between RSA and RSB at the cable end at the data evaluation unit are recommended. 
The following drawing is only designed to illustrate the screening using an example.  
 
 
 
Figure 5.8: Example of RS485 connection (diagram) 
 
In addition to the wiring, various enSuite settings are required before this interface 
can be used.  8.4.5 Changing the IP address basic network and I/O settings”.