114                           Keysight E4980A/AL Precision LCR Meter 
Configuring Measurement Conditions (Display and Function Related Settings)
CORRECTION page
4-
The correction functions of the E4980A/AL are operated as follows:
On/off setting in OPEN, 
SHORT and LOAD fields
OFF ON
Open/short correction at all 
frequency points
Not 
Performed
Open/short correction at 51 preset 
frequency points is performed. 
Correction at other frequency points is 
performed by using interpolation based 
on the correction data at the preset 
frequencies adjacent to the test 
frequency.
Open/short/load correction 
at user-specified spot 
frequency points
Not 
Performed
Open/short/load correction at 
user-specified spot frequency points is 
performed in place of open/short 
correction at all frequency points when 
test frequency is identical with one of 
the user-specified spot frequency 
points.