EasyManua.ls Logo

Keysight E4980AL - System:error[:Next]; System:klock; System:persona:manufacturer[:Name]; System:persona:manufacturer:default

Keysight E4980AL
529 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
376 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
:SYSTem:ERRor[:NEXT]?
Syntax :SYSTem:ERRor[:NEXT]?
Description Returns an error message. (Query Only)
Equivalent key No equivalent key is available on the front panel.
:SYSTem:KLOCk
Syntax :SYSTem:KLOCk {ON|OFF|1|0}
:SYSTem:KLOCk?
Description Enables the front panel key lock. If the key lock’s pin on the handle interface is
set to “LOCK,” touching the keys will have no effect.
Parameter
Equivalent key [Local/Lock]
:SYSTem:PERSona:MANufacturer[:NAME]
Syntax :SYSTem:PERSona:MANufacturer[:NAME] <string>
Description This command only accepts two non-case sensitive strings “Agilent
Technologies” and “Keysight Technologies”. A power cycle or reboot is required
for the changes in the instrument’s manufacturer string to take effect and to be
recognized in the Keysight Connection Expert. The string will remain for the
subsequent power cycle or reboot.
Equivalent key No equivalent key is available on the front panel.
:SYSTem:PERSona:MANufacturer:DEFault
Syntax :SYSTem:PERSona:MANufacturer:DEFault
Description This command sets the instruments manufacturer to “Keysight Technologies
which is the default manufacturer setting.
Equivalent key No equivalent key is available on the front panel.
Description
ON or 1 Enables the Front panel key lock
OFF or 0 (Preset value) Disables the Front panel key lock

Table of Contents