186 Keysight E4980A/AL Precision LCR Meter
Save/Recall
Saving Measurement Results into USB Memory
6-
Figure 6-5 Saving measurement results into USB memory
How to save the measurement result of List Sweep Measurement to USB
memory.
Step 1. Set the List Sweep. For the setting, refer to the “LIST SWEEP SETUP Page” on
page 147.
Step 2. Press [Meas Setup].
Step 3. Using the cursor key, select the TRIG field.
Step 4. Set the trigger mode to (MAN).
Step 5. Set the USB memory to the Front USB Port.
Step 6. Press [Save/Recall].
Step 7. Press soft key SAVE DATA.
Step 8. Press soft key START LOG, then set the mode to accumulate the
measurement result to the data buffer memory.
Step 9. Press [Display Format].
Step 10. Press soft key LIST SWEEP.
Measurement result files are automatically assigned file names
E498x001.csv through E498x999.csv, and you cannot change the file
names.
If the total size of the measurement result data exceeds the capacity of the
data buffer memory, a “Data Buffer Overflow” error message is displayed.
If this happens, press the SAVE & STOP softkey.