Keysight E4980A/AL User’s Guide -491
Handler Interface
Using the Handler Interface
-
Setting Up the Handler Interface Board for Use with the List Sweep
Comparator
Follow these steps to set up the handler interface for the list sweep
comparator:
Step 1. [Meas Setup] - LIST SETUP to open the LIST SWEEP SETUP page.
Step 2. Set up the list table (by defining the sweep points and upper/lower limits) so
that you can conduct list sweep measurement. For more information, see the
chapter on “Configuring Measurement Conditions.”
Step 3. Press the [System] key to open the SYSTEM INFO page.
Step 4. Using the cursor movement keys on the front panel, move the cursor to the
HANDLER I/F field. The following softkeys appear on the softkey label area.
— ON
— OFF
Step 5. Press the ON softkey. The handler interface is enabled for signal
input/output.
Step 6. Press the [Display Format] key to open the MEAS DISPLAY page. Press the
LIST SWEEP softkey so that you can begin list sweep measurement.