372 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
:SYSTem:COMMunicate:LAN[:SELF]:CONFigure
Syntax :SYSTem:COMMunicate:LAN[:SELF]:CONFigure {AUTO|MANual}
:SYSTem:COMMunicate:LAN[:SELF]:CONFigure?
Description Selects automatic obtaining or manual setting for the IP address obtaining
method.
Parameter
Equivalent key [System] - SYSTEM CONFIG - IP CONFIG - AUTO
[System] - SYSTEM CONFIG - IP CONFIG - MAN
:SYSTem:COMMunicate:LAN[:SELF]:CONTrol
Syntax :SYSTem:COMMunicate:LAN[:SELF]:CONTrol
Description Returns the SOCKET control port number. If the parser is a SOCKET, it will
return a number from 5000 to 5100. Otherwise, 0. (Query Only)
Equivalent key No equivalent key is available on the front panel.
:SYSTem:COMMunicate:LAN[:SELF]:CURRent:ADDRess?
Syntax :SYSTem:COMMunicate:LAN[:SELF]:CURRent:ADDRess?
Description Returns the current IP address. (Query Only)
Equivalent key No equivalent key is available on the front panel.
:SYSTem:COMMunicate:LAN[:SELF]:CURRent:DGATeway?
Syntax :SYSTem:COMMunicate:LAN[:SELF]:CURRent:DGATeway?
Description Returns the current Gateway address. (Query Only)
Equivalent key No equivalent key is available on the front panel.
This command is available for the firmware revision 2.00 or later.
Description
AUTO (Preset value) Sets the IP address obtaining method to AUTO
MANual Sets the IP address obtaining method to MAN