358 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
[Meas Setup] - LIST SETUP - LOW
[Meas Setup] - LIST SETUP - HIGH
:LIST:BIAS:CURRent
Syntax :LIST:BIAS:CURRent <point 1>,<point 2>,.... ,<point n>
:LIST:BIAS:CURRent?
Description Clears the previous list sweep table, sets DC current sweep as the list sweep
parameter, and sets DC current points for the DC current list sweep. If the list
sweep is not set for DC current, the query command returns error No. 230. The
value 9.9E37 should be set for any empty points.
Parameter
Equivalent key [Meas Setup] - LIST SETUP - BIAS[A]
:LIST:BIAS:VOLTage
Syntax :LIST:BIAS:VOLTage <point 1>,<point 2>,.... ,<point n>
:LIST:BIAS:VOLTage?
Description Clears the previous list sweep table, sets DC voltage sweep as the list sweep
parameter, and sets DC voltage points for the DC voltage list sweep. If the list
sweep is not set for DC voltage, the query command returns error No. 230. The
value 9.9E37 should be set for any empty points.
Parameter
Equivalent key [Meas Setup] - LIST SETUP - BIAS[V]
:LIST:CLEar:ALL
Syntax :LIST:CLEar:ALL
<point n>
Range of n 1 to 201
Unit A
<point n>
Range of n 1 to 201
Unit V