Keysight E4980A/AL Precision LCR Meter 359
SCPI Command Reference
E4980A/AL commands
10-
Description Clears the list sweep setup. (No Query)
Equivalent key [Meas Setup] - LIST SETUP - No. - CLEAR TABLE
:LIST:CURRent
Syntax :LIST:CURRent <point 1>,<point 2>,... ,<point n>
:LIST:CURRent?
Description Clears the previous list sweep table, sets AC current sweep as the list sweep
parameter, and sets AC current points for the AC current list sweep. If the list
sweep is not set for AC current, the query command returns error No. 230. The
value 9.9E37 should be set for any empty points.
Parameter
Example of Use
viVPrintf(AgtE4980A/AL, ":LIST:CURR 0.01, 9.9e37, 0.02" +
vbLf, 0)
Equivalent key [Meas Setup] - LIST SETUP - LEVEL[A]
:LIST:DCSource:VOLTage
Syntax :LIST:DCSource:VOLTage <point 1>,<point 2>,... ,<point n>
:LIST:DCSource:VOLTage?
Description Clears the previous list sweep table, sets DC source voltage sweep as the list
sweep parameter, and sets DC source voltage points for the DC source voltage
list sweep. If the list sweep is not set for DC source voltage, the query
command returns error No. 230. The value 9.9E37 should be set for any empty
points.
Parameter
Equivalent key [Meas Setup] - LIST SETUP - DC SRC[V]
<point n>
Range of n 1 to 201
Unit A
<point n>
Range of n 1 to 201
Unit A