Keysight E4980A/AL Precision LCR Meter                                                                                                199
Measurement Procedure and Examples
Four-Terminal Pair Configuration
7-
Four-Terminal Pair Configuration
Generally, any mutual inductance, interference of the test signals, or unwanted 
residual factors in the connection method incidental to ordinary termination 
methods will have a significant effect on measurement, especially at high 
frequency. The E4980A/AL employs a four-terminal pair measurement 
configuration that permits easy, stable, and accurate measurement while 
avoiding the negative influences inherent in the above factors. The 4TP 
configuration can improve the impedance measurement range to below 1 m
Figure 7-7 shows the four-terminal pair measurement principle. The set of 
UNKNOWN terminals consists of four coaxial connectors.
• HCUR: High current
• HPOT: High potential
• LPOT: Low potential
• LCUR: Low current
Figure 7-7  Four-terminal pair measurement principle
Figure 7-7 *1 The shielding conductor of Lcur is shunted to the system ground 
in the E4980A/AL
The four-terminal pair (4TP) configuration solves the effects of mutual coupling 
between the leads by employing either or both of the following techniques. The 
E4980A/AL employs the method 2. At frequencies higher than 10kHz, the 
effects of 1. is also benefited on the E4980A/AL by the balun effects of the 
coaxial test leads
1. The outer shield conductors work as the return path for the test signal 
current (they are not grounded). The magnetic fields produced by the inner 
and outer currents cancel each other because of the opposite directions 