344 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
Parameter
Equivalent key [Display Format] - DISPLAY BLANK
:DISPlay:LINE
Syntax :DISPlay:LINE <String>
:DISPlay:LINE?
Description Enters arbitrary comments containing up to 30 ASCII characters in the
comment field. The string “USER COMMENT” is displayed if this is empty.
Parameter
Equivalent key [Meas Setup] - MEAS SETUP - USER COMMENT - ADD CHAR
[Meas Setup] - MEAS SETUP - USER COMMENT - ENTER
[Meas Setup] - MEAS SETUP - USER COMMENT - NEXT
[Meas Setup] - MEAS SETUP - USER COMMENT - PREV
:DISPlay:PAGE
Syntax :DISPlay:PAGE
{MEASurement|BNUMber|BCOunt|LIST|MSETup|CSETup|LTABle|LSETup|CATAlo
g|SYSTem|SELF|MLARge|SCONfig|SERVice}
:DISPlay:PAGE?
Description Selects the page to be displayed.
Parameter
Description
ON or 1 (Preset value) Enables display updates
OFF or 0 Disables display updates. The display will be blank.
<String>
Preset value “” (NULL)
Description
MEASurement (Preset value) Sets displayed page to <MEAS DISPLAY>
BNUMber Sets displayed page to <BIN No. DISPLAY>