Keysight E4980A/AL Precision LCR Meter 339
SCPI Command Reference
E4980A/AL commands
10-
Equivalent key [Meas Setup] - CORRECTION - SHORT - ON
[Meas Setup] - CORRECTION - SHORT - OFF
:CORRection:SPOT[1-201]:FREQuency
Syntax :CORRection:SPOT[1-201]:FREQuency <numeric>
:CORRection:SPOT[1-201]:FREQuency?
Description Sets the frequency for the specified measurement point. The mode selected by
:CORRection:METHod is chosen as the correction mode for the specified point.
Parameter
Equivalent key [Meas Setup] - CORRECTION - FREQ
:CORRection:SPOT[1-201]:LOAD[:EXECute]
Syntax :CORRection:SPOT[1-201]:LOAD[:EXECute]
Description Executes LOAD correction at the specified measurement point. (No Query)
Equivalent key [Meas Setup] - CORRECTION - FREQ - MEAS LOAD
:CORRection:SPOT[1-201]:LOAD:STANdard
Syntax :CORRection:SPOT[1-201]:LOAD:STANdard <reference value for primary>,
<reference value for secondary>
:CORRection:SPOT[1-201]:LOAD:STANdard?
Description Sets the standard reference values at the specified measurement point. The
measurement function for the reference values can be selected using
:CORRection:LOAD:TYPE.
OFF or 0 (Preset value) Disables SHORT correction
Description
<Numeric>
Range 20 to 2M
Preset value 20
Unit Hz