352 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
Equivalent key [Meas Setup] - MEAS SETUP - REF A
[Meas Setup] - MEAS SETUP - REF B
:FUNCtion:IMPedance:RANGe:AUTO
Syntax :FUNCtion:IMPedance:RANGe:AUTO {ON|OFF|1|0}
:FUNCtion:IMPedance:RANGe:AUTO?
Description Enables the auto-range function for impedance measurement.
Parameter
Equivalent key [Display Format] - MEAS DISPLAY - RANGE - AUTO
[Display Format] - MEAS DISPLAY - RANGE - HOLD
[Meas Setup] - MEAS SETUP - RANGE - AUTO
[Meas Setup] - MEAS SETUP - RANGE - HOLD
:FUNCtion:IMPedance:RANGe[:VALue]
Syntax :FUNCtion:IMPedance:RANGe[:VALue] <numeric>
:FUNCtion:IMPedance:RANGe[:VALue]?
Description Selects the impedance measurement range. This command turns the auto
range function OFF.
Parameter
Unit Depends on
:FUNCtion:IMPedance[:TYPE].
<Numeric>
Description
ON or 1 (Preset value) Auto range ON
OFF or 0 Auto range OFF
<Numeric>
Range 100m|1|10|100|300|1k|3k|10k|30k|100k
Preset value 100
Unit ohms