338 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
:CORRection:OPEN[:EXECute]
Syntax :CORRection:OPEN[:EXECute]
Description Executes OPEN correction based on all frequency points. (No Query)
Equivalent key [Meas Setup] - CORRECTION - OPEN - MEAS OPEN
:CORRection:OPEN:STATe
Syntax :CORRection:OPEN:STATe {ON|OFF|1|0}
:CORRection:OPEN:STATe?
Description Enables OPEN correction.
Parameter
Equivalent key [Meas Setup] - CORRECTION - OPEN - ON
[Meas Setup] - CORRECTION - OPEN - OFF
:CORRection:SHORt[:EXECute]
Syntax :CORRection:SHORt[:EXECute]
Description Executes SHORT correction based on all frequency points. (No Query)
Equivalent key [Meas Setup] - CORRECTION - SHORT - MEAS SHORT
:CORRection:SHORt:STATe
Syntax :CORRection:SHORt:STATe {ON|OFF|1|0}
:CORRection:SHORt:STATe?
Description Enables SHORT correction.
Parameter
Description
ON or 1 Enables OPEN correction.
OFF or 0 (Preset value) Disables OPEN correction.
Description
ON or 1 Enables SHORT correction