Keysight E4980A/AL Precision LCR Meter 363
SCPI Command Reference
E4980A/AL commands
10-
Example of Use viVPrintf(AgtE4980A/AL, "LIST:STIM:TYPE FREQ, VOLT" + vbLf, 0)
viVPrintf(AgtE4980A/AL, "LIST:STIM:DATA 1E3, 1, 2E3, 1.1, 3E3,
0.9" + vbLf, 0)
Equivalent key No equivalent key is available on the front panel.
:LIST:VOLTage
Syntax :LIST:VOLTage <point 1>,<point 2>,... ,<point n>
:LIST:VOLTage?
Description Clears the previous list sweep table, sets AC voltage sweep as the list sweep
parameter, and sets AC voltage points for the AC voltage list sweep. If the list
sweep is not set for AC voltage , the query command returns error No. 230. The
value 9.9E37 should be set for any empty points.
Parameter
Example of Use
viVPrintf(AgtE4980A/AL, ":LIST:VOLT 0.5, 0.8, 1, 9.9e37, 1.5"
+ vbLf, 0)
Equivalent key [Meas Setup] - LIST SETUP - LEVEL[V]
:MEMory:CLEar
Syntax :MEMory:CLEar DBUF
FREQuency Selects frequency as the second stimulus parameter
VOLTage Selects voltage as the second stimulus parameter
CURRent Selects current as the second stimulus parameter
BVOLtage Selects bias voltage as the second stimulus parameter
BCURrent Selects bias current as the second stimulus parameter
DCSVoltage Selects DC source voltage as the second stimulus parameter
Second Parameter
<point n>
Range of n 1 to 201
Unit V