Keysight E4980A/AL Precision LCR Meter 357
SCPI Command Reference
E4980A/AL commands
10-
Equivalent key No equivalent key is available on the front panel.
:INITiate[:IMMediate]
Syntax :INITiate[:IMMediate]
Description Initiates the trigger to change from the Idle” state to the “Wait for Trigger”
state one time. Refer to the chapter on Remote Control. (No Query)
Equivalent key No equivalent key is available on the front panel.
:LIST:BAND[1-201]
Syntax :LIST:BAND[1-201] {A|B|OFF},<low>,<high>
:LIST:BAND[1-201]?
Description Sets the limit values of the limit function for the list sweep measurement. When
A or B is the parameter, both lower and the upper limit should be input. If
either of them is not input, error No. 109 is returned. In the case of
OFF,<low>,<high> are optional.
Parameter
Equivalent key [Meas Setup] - LIST SETUP - LMT - A
[Meas Setup] - LIST SETUP - LMT - B
[Meas Setup] - LIST SETUP - LMT -
OFF or 0 (Preset value) Disables automatic trigger state change
Description
Description
A Sets limit test target to A (Primary parameter)
B Sets limit test target to B (Secondary parameter)
OFF (Preset value) Sets limit test target to OFF
<low>,<high>
Preset value Query returns -9.9E37, +9.9E37 for <low>, <high>,
respectively in the case of OFF.
Unit Depends on :FUNCtion:IMPedance[:TYPE]