Keysight E4980A/AL Precision LCR Meter 379
SCPI Command Reference
E4980A/AL commands
10-
Parameter
Equivalent key [Meas Setup] - MEAS SETUP - TRIG - INT
[Meas Setup] - MEAS SETUP - TRIG - MAN
[Meas Setup] - MEAS SETUP - TRIG - EXT
[Meas Setup] - MEAS SETUP - TRIG - BUS
:TRIGger:TDEL
Syntax :TRIGger:TDEL <numeric>
:TRIGger:TDEL?
Description Sets the trigger delay time.
Parameter
Equivalent key [Meas Setup] - TRIG SETUP - TRIG DLY
:VOLTage[:LEVel]
Syntax :VOLTage[:LEVel] <numeric>
:VOLTage[:LEVel]?
Description Sets the voltage level for the measurement signal. If the signal is set by
:CURRent[:LEVel], the query command returns error No. 230.
Description
INTernal (Preset value) Sets trigger source to “internal”
HOLD Sets trigger source to “manual”
EXTernal Sets trigger source to “external connector on the rear
panel”
BUS Sets trigger source to “GPIB/LAN/USB”
<Numeric>
Range 0 to 999
Preset value 0
Unit s
Resolution 100