Keysight E4980A/AL Precision LCR Meter 341
SCPI Command Reference
E4980A/AL commands
10-
:CORRection:USE[:CHANnel]
Syntax :CORRection:USE[:CHANnel] <channel>
:CORRection:USE[:CHANnel]?
Description Selects the channel number to be used for the Scanner I/F MULTI mode.
Parameter
Equivalent key [Meas Setup] - CORRECTION - CH
:CORRection:USE:DATA[:MULTi]
Syntax :CORRection:USE:DATA[:MULTi] <channel>,<open1 A>,<open1 B>,<short1
A>,<short1 B>,<load1 A>,<load1 B>,<open2 A>,<open2 B>,<short2 A>,<short2
B>,<load2 A>,<load2 B>,.......,<open 201A>,<open201 B>,<short201
A>,<short201 B>,<load201 A>,<load201 B>
:CORRection:USE:DATA[:MULTi]? <channel>
Description Sets or returns the open/short/load data for the “Multi” correction mode. The
data for all 201 test points must be set. Any point with no correction data
should be set to 0. A query will return 1,206 data for 201 test points even if
there are some frequency points that were turned off. A value of 0 is returned
when there are the no correction data.
Parameter
Example of Use
Dim Result As String * 50000
<channel>
Range 0 to 127
Preset value 0
<channel>
Range 0 to 127
<open(n) A>, <open(n) B>, <short(n) A>, <short(n)
B>, <load(n)A>, <load(n) B>
Preset value 0
Units OPEN: S; SHORT: ohms; LOAD: Depends on
:CORRection:LOAD:TYPE