360 Keysight E4980A/AL Precision LCR Meter
SCPI Command Reference
E4980A/AL commands
10-
:LIST:FREQuency
Syntax :LIST:FREQuency <point 1>,<point 2>,... ,<point n>
:LIST:FREQuency?
Description Clears the previous list sweep table, sets frequency sweep as the list sweep
parameter, and sets frequency points for the frequency list sweep. If the list
sweep is not set for frequency, the query command returns error No. 230. The
value of 9.9E37 should be set for any empty points.
Parameter
Example of Use
viVPrintf(AgtE4980A/AL, ":LIST:FREQ 1e2,2e2,5e2,9.9e37,1e3" +
vbLf, 0)
Equivalent key [Meas Setup] - LIST SETUP - FREQ[Hz]
:LIST:MODE
Syntax :LIST:MODE {SEQuence|STEPped}
:LIST:MODE?
Description Selects the sweep mode for the list sweep measurement function. In the case
of SEQ, one trigger makes all sweep point measurements. In the case of STEP,
one trigger makes single point-by-point measurements.
Parameter
Equivalent key [Meas Setup] - LIST SETUP - MODE - SEQ
[Meas Setup] - LIST SETUP - MODE - STEP
:LIST:SEQuence:TSTamp:CLEar
Syntax :LIST:SEQuence:TSTamp:CLEar
<point n>
Range of n 1 to 201
Unit Hz
Description
SEQuence (Preset value) Sets the list sweep mode to sequence mode
STEPped Sets the list sweep mode to step mode